
19.1 FUNCTION OF DIGITAL INPUTS
Meter Reset Logical 0 -The microcomputer reads and resets the meter to the values
stored in nonvolatile memory. If totals are saved on power down,
totals are reset to the saved value, otherwise totals are set to zero
or to the offset value if offset is not zero.
Function Reset Logical 0 -All totals are reset to zero or to the offset value if offset is not zero,
alarms are reset and peak display resets to minus overload
Peak Display Logical 0 - The peak value of the filterable item is displayed
Logical 1 - The present value the selected item is displayed.
Hold Logical 0 - The meter display and outputs are held at the last reading prior to
the hold going low. Meter continues to totalize while meter is in
hold
Logical 1 -The display and outputs are updated normally
External Gate Logical 0 -The gate time starts when input goes low and ends when gate
time goes high
Logical 1 -When external gate is selected, it overrides the internal gate. At
power on, the meter displays reset until the external gate input
goes low and then high to complete a conversion. The new value
will then be displayed until a new external gate occurs.
Display Blank Logical 0 - The display and indicator lights are blanked
Logical 1 - The display and indicator lights are lit.
External Gate Logical 0 - Gate Time is closed
Logical 1 - Gate time is open and remains open until logical 0 applied
Display Item 2 & Display Item 3
Inputs A & B Logical 0 or 1 - Item 1 displayed
Input A Logical 0 and Input B Logical 1 - Item 2 displayed
Input B Logical 0 and Input A Logical 1 - Item 3 displayed
19. DIGITAL INPUTS
-47-
Kommentare zu diesen Handbüchern