In batch control mode with “external gate”, the meter waits a the end of every cycle
until an external gate input is grounded for a minimum of 3.33 ms. This starts a new fill
cycle by energizing Relay #1 and resetting the Batch Total. Gate time is not used.
Three values are tracked and can be separately displayed by pressing the RESET key: Item
#1, the Batch Total; Item #2, the Grand Total of all batches or Number of Batches (select-
able during setup); and Item #3, the Fill Rate.
Item #1, Batch Total, is the total of input pulses for that batch. It may be configured to
count up from 0 to a preset, or to count down from a preset to 0. The preset value is
placed in ALARM1. SCALE1 is positive whether counting up or down.
Item #2, Grand Total, is the sum of previous Batch Totals and the current Batch Total. It
can overflow to exponential format.
Item #2 (alternate), Number of Batches, is the current count of batches. SCALE1 does
not apply. dECPt1 is set to 1.
Item #3, Fill Rate, is calculated with a fixed 20 ms gate time. It may be displayed as Item
#3.
BATCH CONTROL WITH A 4-20 mA OUTPUT FLOW METER
Application:
Fill 55 gallon tanks. Use a
4-
20 mA flow meter where 4 mA = 0 and 20
mA = 39.20 GPM. Slow down filling at 54
gallons. Cycle batches automatically with 20
sec between cycles. Displ
ay batch total & fill
rate to 2 places. Also track num
ber of
batches.
Solution:
Use an Extended counter with
a dual relay output board. Set Input to “Rate
Batch.” Set Batch to count up to ALARM1, to
use Gate Time as delay between batches,
and to make It
em #2 the number of batches.
Set Gate Time to 20 sec. Set DecPt1 and
DecPt 2 to two decimal places for Items #1
and #3 (Batch Total and Rate). Scale Item
#3 (Rate) using the co
ordinates of 2 points
method so that 20.0000 mA will be dis
played
as 39.20 GPM
. Scale Item #1 (Batch Total)
by entering a Scale1 of 1.66667 and a mul-
tiplier of 0.01. That is because totalizing
sums read
ings in gallons every second.
Since our rate is in GPM, we have to divide
by 60. E
nter an Offset1 of 55.00to serve as
the batch se
tpoint in gallons. Set Setpoint2
to 54.00 to activate Relay 2 to slow the fill
rate.